Devices
Devices
Devices (in the Probes menu) lists every probe at the sites you can see, whether each one is reporting, and when it is due for calibration. It is also where a probe's limits and alarm settings are set.
How probes get here
- SmartProbe adds every probe to your account and allocates it to a site. A new probe appears here once it has been allocated to you.
- Serial numbers, install dates and calibration dates are set by SmartProbe. You'll find them under Diagnostics on the probe's page.
- The list shows Last Sync, Location and Calibration due for each probe. Filter narrows it, for example to probes overdue for calibration or low on battery. Filters are remembered, so use Clear filters if the list looks empty.
Naming and limits
Masters change these with Edit, or Edit Thresholds for just the limits. Other roles can view them.
- Label: what your team calls the probe, such as "Walk-in Cooler". It appears in alerts and reports.
- Each channel has Warning Low, Alert Low, Warning High and Alert High limits:
- Past a warning limit: a warning incident on the dashboard only. No email or SMS.
- Past an alert limit: a critical incident, with SMS and email alerts once it has lasted longer than the initial alarm delay.
- Keep warning limits inside the alert limits.
- A channel with no limits is monitor-only: its readings are recorded, but it never raises an alert.
- Initial Alarm Threshold, Alarm Interval and Offline Threshold use your workspace defaults unless you enter a number for this probe.
- Turn on Process probe so the probe is offered when starting a cook chill batch or a spot check.
Saved changes reach the site's tablet on its next check-in.
Calibration certificates
- SmartProbe calibrates your probes, tracks every due date, and adds each certificate to the probe's Calibration History, with View Certificate and Download PDF.
- Calibration due is red when overdue, amber within 30 days, and Never calibrated when there is no certificate. The Probe care figures on the dashboard open this list filtered.
- For an auditor, select probes and use Download Calibration Certificates to get a ZIP of each probe's latest certificate.
- Every probe log report states the certificates in force over its time window.
Moving a probe
- To move a probe to another of your sites, a Master changes its Location on Edit.
- The probe keeps its history. Its past readings stay with the site they were taken at, in that site's time zone.
Offline and battery
- If a probe sends nothing for longer than its Offline Threshold, an offline incident is raised. If a whole site goes quiet at once, it is one Tablet Not Syncing incident for the site.
- Offline problems are only detected once someone is subscribed to Device Offline/No Signal or Tablet Not Syncing in Notification Manager.
- If a probe stops reporting, check the SmartProbe tablet at the site first.
- Turn off Expected to report constantly for probes used only now and then, such as cook chill probes, so nobody is alerted when they go quiet.
- Low battery probes show on the dashboard's Probe care figures and raise a Low Battery incident.
Operating hours
For equipment that is genuinely switched off at set times, turn on Enable operating-hours timeclock and add a window for each day it runs. Readings outside those hours are withheld from logs, reports, exports and alarms, so use it only when the equipment is really off.